  • ECG Daily Maintenance
    Battery Maintenance      
    It is required to charge & discharge the battery at least every 3 months.      
    When charging, you are required to keep the machine in standby condition.      
    Indicator on the top the unit will blink while charging, and keep in light when charging is   
    almost completed that it requires additional 2~3 hours to complete the charging.    
    Thermal Printer Maintenance  Residue and dirt on the thermal printer could affect the clarity of printing out ECG trace.     
    To clean the thermal printer, you are required to open the paper magazine and clean the   
    printer  with soft cotton dipped with alcohol.      
    It is not permitted to operate on the printer with a sharp object.      
    Otherwise, permanent damage could be resulted.      
    Thermal printer maintenance should be done at least once a month
  • Printer Failure
    Printer failure covers such circumstance that there will be some vertical break in the 
     printed trace.Whenever a printer fault occurs, you are required to clean the thermal printer with  
    soft cotton with alcohol. If this action does not work, certain thermal emitting component is  
    probably damaged. If printing out trace is blurring, you are required to do as follows:  
    ●Open the paper magazine.  
    ●Adjust the screw located below the printers shown in the following figure.
  • Lead Failure
    Lead fault covers the situation that some lead does not give trace in the process of recording that 
    is probably caused by the following. If you start recording immediately after the patient cable is 
    connected to the patient, the unit will probably face problem of lead fault as a result of saturation 
    or overflow at the baseline adjustment software due to receiving and working on unstable ECG 
    trace & data. In such case, provided good lead connection is guaranteed, you’d better press key 
    “RESET” to rapidly stabilize the trace of each channel or to try another recording, and everything 
    will be OK. Damaged patient cable may also cause problem of lead fault. TO do checks on the 
    patient cable, and contact our service technicians if it is damaged. The last reason for lead fault is
     the unit problem due to failure in signal communication. If you are sure to exclude other cause for
     this problem, please do not hesitate to contact our service technicians.
  • Frequently Asked Questions of Ultrasonic Scanner
    Good image but the mouse doesn't work well:
    a) Refer to How to Clean the Mouse  
    Image isn't clear enough:
    a) Adjust the far, medium or near field and the gain; 
    b) Adjust the brightness and contrast; 
    c) Clean the organic glass of the display;  
    Interrupted Stripe and snow-shaped interference on the screen:
    a) Check if it power supply is interfered by other electrical equipment;  
    b) Check if operation condition is free from any electrical or magnetic interference;  
    c) Check if power supply, probe and plugs are well connected or not;  
    d) Far field interference is strong when you hold the probe in hand but it turns ok when you put it down; 
    Please find out the interference source; 
    e) Check if the grounding wire is connected or not;  
    Characters are displayed on the screen but no images available when the power supply is switched on:
    a) Check if the plug of the probe is correctly connected, 
    b) Check if the Near Field, Far Field and Gain are correctly adjusted, ensure that they are not turned to the Min.value. 


    The indictor is still off and no display on the screen after the unit is switched on;
    1) Check power supply 
    2) Check the power cord and plug 
    3) Check if the fuse is broken; 
    4) Turn the knobs "BRIGHT" and "CONTRAST"  